Home » Abstracts » Technological & Instrument advances in MS » Enhanced Immunopeptide Identification Using MIDIA-PASEF: A Novel timsTOF Scan Mode

Enhanced Immunopeptide Identification Using MIDIA-PASEF: A Novel timsTOF Scan Mode

Author(s)

  • Jonathan Krieger | Bruker Dalotnik GmbH & Co KG | Fahrenheitstrasse 4, 28359, Bremen, Germany
  • Diego Assis | Bruker Daltonics | 40 Manning Road, MA 01821, Billerica, United States
  • Matthew Willets | Bruker Daltonics | 40 Manning Road, MA 01821, Billerica, France
  • Pierre-Olivier Schmit (Presenting Author) | Bruker France SAS | 34, rue de l'industrie, 67160, Wissembourg, France
  • Nicholas Cheung | Regeneron Pharmaceuticals, Inc | 777 Old Saw Mill River Road, NY 10591-06707, Tarrytown, United States
  • Robert Salzler | Regeneron Pharmaceuticals, Inc | 777 Old Saw Mill River Road, NY 10591-06707, Tarrytown, United States

Abstract

Mass Spectrometry (MS)-based Immunopeptidomics (Ipep) provides biologically relevant information that cannot be obtained through other techniques. While Data-Independent Acquisition (DIA) methods tend to be more reproducible and sensitive than Data Dependent Acquisition (DDA) methods, the resulting data can be challenging to analyze. Recently, a new acquisition method, midia-PASEF (Distler, 2023), has shown promise for Ipep by acquiring data in the DIA space while processing it in a DDA manner.
Immunopeptides were enriched from varying amounts of B-lymphocytes (from 1x10^6 to 5x10^6 cells) and analyzed using PASEF and midia-PASEF with a timsTOF Ultra 2 system (Bruker). Both acquisition schemes employed ion mobility selection to isolate peptides of interest (+1 for m/z > 600, +2, +3) while minimizing background noise. The resulting data were analyzed with the open-source Sage search engine, combined with MS2Rescore (used as part of the midiaID processing pipeline for midia-PASEF data).
Regarding Class I immunopeptides, we observed a 20% increase of identified peptides identified using midia-PASEF. At each loading level, the amino acid motifs for the 9-mers were similar between both methods, including the preference for E or L at position 2 and V,L,I or A at position 9. This suggests that the additional peptides identified using midia-PASEF are relevant. By combining de novo sequencing with midia-PASEF acquisition, we were able to identify 10% more immunopeptides.