Home » Abstracts » Technological & Instrument advances in MS » Design and preliminary evaluation of a microfluidic device to automate and improve high-throughput sample preparation for low cells number to single-cell proteomics

Design and preliminary evaluation of a microfluidic device to automate and improve high-throughput sample preparation for low cells number to single-cell proteomics

Author(s)

  • Christopher Kune | Mass Spectrometry Laboratory – Molsys research unit, University of Liège | Allée du six aout, 11 Building B6c Quartier Agora, 4000, Liège, Belgium
  • Ali Gholizadeh | Microfluidic Lab, University of Liège | Bât. B52/3, Quartier Polytech, 1 allée de la Découverte 9, 4000, Liège, Belgium
  • Sylvia Tielen | GIGA Stem Cells - Molecular Regulation of Neurogenesis, University of Liège | Quartier Hôpital 15, Avenue Hippocrate, 4000, Liège, Belgium
  • Denis Vandormael | Sirris | Liège Science Park, Rue du Bois Saint-Jean 12, 4102, Seraing, Belgium
  • Maximilien Fléron (Presenting Author) | GIGA Proteomics Facility, University of Liège | Allée du six aout, 11 Building B6c Quartier Agora, 4000, Liège, Belgium
  • Dominique Baiwir | GIGA Proteomics Facility, University of Liège | Allée du six aout, 11 Building B6c Quartier Agora, 4000, Liège, Belgium
  • Laurent Nguyen | GIGA Stem Cells - Molecular Regulation of Neurogenesis, University of Liège | Quartier Hôpital 15, Avenue Hippocrate, 4000, Liège, Belgium
  • Tristan Gilet | Microfluidic Lab, University of Liège | Bât. B52/3, Quartier Polytech, 1 allée de la Découverte 9, 4000, Liège, Belgium
  • Gauthier Eppe | Mass Spectrometry Laboratory – Molsys research unit, University of Liège | Allée du six aout, 11 Building B6c Quartier Agora, 4000, Liège, Belgium
  • Gabriel Mazzucchelli | Mass Spectrometry Laboratory – Molsys research unit, University of Liège | Allée du six aout, 11 Building B6c Quartier Agora, 4000, Liège, Belgium

Abstract

Current LC-MS systems allow to analyze proteomes from cell populations to single cell. However, reducing the starting sample material can compromise performance, making effective sample preparation essential. A major challenge is peptides loss due to surface adsorption.
Peptides intensity decreases with their concentration, even when the same quantity is injected onto the LC column. To minimize peptides adsorption, we recommend to reduce sample volume to minimize exposure and increase concentration, avoid sample transfers, and use materials with low peptide affinity. Sample preparation throughput is also a challenge, as thousands of samples are analyzed for a single-cell proteomics study.
To address these challenges, we developed a microfluidic device that automates all sample preparation steps using centrifugation, from cell collection to LC-MS injection, at a sub-microliter scale. Proteomic sample preparation is performed by successively adding 15 nL of reagent for protein reduction, alkylation and digestion, followed by releasing the sample into an LC-compatible 384-well plate. The design allows for serial processing, increasing throughput, and includes sample multiplexing options. Custom vials compatible with cell sorting and LC/MS instruments, molded from low-affinity materials are also evaluated.
Preliminary fluidic results of the microfluidic device and advantages of the one-pot strategy applied with adequate polymeric materials will be presented.