Abstract
Plasma proteomics is promising for discovering disease biomarkers. However, large-scale peptide PRM requires a robust MS setup without compromising protein identification, dynamic range or precision. This necessitates a scheduled retention time (RT) approach, which could cause missed peaks. Rescheduling RT windows can be tedious.The adaptive RT on a modified Thermo Scientific Orbitrap hybrid MS can solve this problem. This method dynamically adjusts scheduled acquisition windows at narrower time intervals eliminating missing values. This is the first time adaptive RT has been used on Orbitrap Hybrid MS in large-scale targeted assays.
Methods
PQ500 peptides (Biognosys) were spiked into 400ng of neat plasma digest (Pierce). Thermo Scientific Vanquish Neo LC in trap&elute mode was used with a PepMap column.
Results
25 replicates were injected and analyzed over a 27 min gradient. Without adaptive RT, RT shifted by 0.5-1.5 min with different column conditions. With adaptive RT, shifted RT windows were captured. No RT rescheduling was required after changing column conditions. The system repetitively quantifies ~800 peptides during a 27-min gradient with different column conditions. Dilution curves of PQ500 in plasma were performed to assess sensitivity. The majority of LODs were less than 50 attomoles, with some LODs less than 10 attomoles. >94% peptides had CV values less than 25%. The standard curves of the spiked peptides showed good linearity across wide dynamic ranges.