Abstract

In recent years, there has been a shift towards analyzing complex proteomics samples more quickly. We optimized a dia-PASEF acquisition mode using narrow isolation windows of 5Da, combined with focused m/z and ion mobility window placements, to evaluate performance on complex proteomics samples.
Tryptic in-house digests of human cell lysate, yeast, and E. coli were used to assess the performance of narrow-window dia-PASEF for short gradients on the timsTOF HT. 800ng of sample digests were loaded on a 25cm C18 column and separated using a 15-minute ACN gradient. Data were processed in Spectronaut (v19, Biognosys) using library-free mode.
Using the narrow-window dia-PASEF method, we measured nearly the complete yeast proteome in 15 minutes, identifying 4677 protein groups from triplicate injections. From a human protein digest sample, representing a higher complexity proteome, we identified nearly 135,000 stripped peptide sequences from 8850 protein groups. Samples mixed in defined ratios were used to evaluate the setup for quantitative proteomics. Within a 15-minute gradient, we identified and quantified 14,366 protein groups and 170,917 peptides. The quantitative performance showed low median coefficient of variation for replicate runs and high quantitative accuracy with low standard deviation, making the narrow-window dia-PASEF method well-suited for investigating proteome-wide changes in protein abundance in various biological contexts.